- Fixed flange
- Manual controlled double-layer shutters
-
Retractable measure tape
-
High brightness, low power consumption LED
-
Support external DAP and additional filters
-
Easy installation
Technical Specifications:
Operation Mode |
Manual control |
Maximum Energy (kV) |
150 |
Al Equivalent, IEC 60601-1-3 |
1 mm Al |
Rectangular X-ray Field Selection @SID=100 cm (cm²) |
Min: 2 × 2, Max: 43 × 43 |
Light Field / X-ray Field Max. Deviation @SID=100 cm (cm²) |
≤ 1% of SID |
Illumination Intensity @SID=100 cm IEC 60601-2-54 |
≥ 180 lux |
Accuracy of Light Field Indicator IEC 60601-2-54 |
≤ 1% of SID |
Light Source On Time (s) |
30 |
Contrast Ratio of Light IEC 60601-2-54 |
≥ 4:1 |
Maximum Radiation Leakage EN 60601-1-3 |
< 1 mGyh |
Power Supply |
24V AC/DC (± 10%), Max. 4A |
Dimensions (mm3 ) |
206 × 230 × 130.5 |
Weight (kg) |
7.0 ± 0.5 |
Focus to Mounting Plate (mm) |
56 |
*Please note: Installation fee does not include travel and accommodation costs for the engineer. To request installation, please fill the order form: CLICK HERE